Characteristics MAVOLUX 5032B
Most Important Error Limits MAVOLUX 5032B
Light Detector silicon photodiode with
V(
λ) filter per DIN 5032 part 7,
class B
Error Limits per DIN 5032, part 7, class B
Meas. Quantity Measuring Range Resolution
in lx
Resolution
in fc
in Lux (lx) in footcandles (fc)
Illumination I 0.01
19,99 0.001 1.999 0.01 0.001
I I 0.1
1 99,9 0.01 19.99 0.1 0.01
III 1
19 99 0,1 1 99.9 1 0.1
I V 10
19 990 1 19 99 10 1
in Candela/m² (cd/m²) in footlambert (fL) in cd/m² in fL
Luminance
(with luminance
attachment
for cd/m
2
or fL)
I0.1
199.9 0.01 19.99 0.1 0.01
II 1
1 999 0.1 199.9 1 0.1
I I I 10
19 990 1 1 999 10 1
I V 100
199 900 10 19 990 100 10
Characteristics Allowable Error
Limits per
DIN 5032 Class B
Actual Error
MAVOLUX 5032B
V(
λ) Matching 6% 3%
True Cosine Evaluation 3% 2%
Linearity 2% 1%
Adjustment Error 1% 0,8%
Total Error 10% 8%