Gossen MAVOSPEC Instrukcja Użytkownika Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 29
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 17
GOSSEN Foto- und Lichtmesstechnik GmbH 18
3 Trigger Input / Output
MAVOSPEC is equipped with input and output signal lines dedicated for starting
measurement from external device and triggering external devices when measurement
is started.
Trigger In line - high state on this line causes the same effect as pressing button on the
device.
Trigger Out line - at the measurement start MAVOSPEC sets high state on this line.
The high state is maintained throughout the period of measurement.
Trigger lines state:
Trigger Technical Data
Connector
Minijack 3.5mm, stereo, 3-pin
Trigger in line
Input line, type open collector with galvanic
isolation. Voltage 4.5 ... 9 V. Min current 1 mA
Trigger in signal
Recommended high state >= 50 µs,
Trigger out line
Output line, type open collector with galvanic
isolation. Voltage 3 ... 9 V.
Recommended current <= 1 mA
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 28 29

Komentarze do niniejszej Instrukcji

Brak uwag